Basic Info.
Model NO.
JYK-801C
Type of Body Fluid-Processing Device
Ingredients Blood Transfusion Device
Volume Alarm
8 Levels
Display
LCD
Transport Package
Carton
Specification
CE
Trademark
JYK
Origin
China
Production Capacity
50000/Month
Product Description
Medical equipment, volumetric infusion pump JYK-801C
The Features of Volumetric Infusion Pump
1. Complies with IEC 60601, higher security for infusion.
2. Volumetric infusion pump, with flow rate accuracy ≤ ± 5%.
3. Unique automatic calibration function to enhance accuracy.
4. Anti-reverse function on motor to prevent upstream.
5. Self-check when start up and during infusion.
6. Pressure and air-in-line detection devices both with 8 sensitivities setting.
7. Adjustable 8 volume levels of alarm for convenient operation in different environments.
8. Up to 10 adjustment values for various brands of IV set could be stored to make the operation more convenient.
9. CE certified.
The Features of Volumetric Infusion Pump
1. Complies with IEC 60601, higher security for infusion.
2. Volumetric infusion pump, with flow rate accuracy ≤ ± 5%.
3. Unique automatic calibration function to enhance accuracy.
4. Anti-reverse function on motor to prevent upstream.
5. Self-check when start up and during infusion.
6. Pressure and air-in-line detection devices both with 8 sensitivities setting.
7. Adjustable 8 volume levels of alarm for convenient operation in different environments.
8. Up to 10 adjustment values for various brands of IV set could be stored to make the operation more convenient.
9. CE certified.
Infusion Mode | Volumetric |
Pumping Mechanism | Peristaltic finger pump |
IV Sets | Standard IV sets of 10, 15, 20 and 60d/mL, with diameter 3.4 ~ 4.5mm. |
FlowRate Range | 1.0 ~ 999.9 mL/h |
Increment | 0.1 mL/h |
Flow Rate Accuracy | ≤ ± 5% (after calibration) |
Volume Limit | 1~9999 mL |
Volume Infused Display | 0~9999 mL |
Purge Rate | 600 mL/h |
KVO Rate | 1 mL/h |
Occlusion Pressure | Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa. |
Air-in-line Detection | Air bubble detected by ultrasonic and the device with 8 sensitivities setting,50 ~ 1000μL. |
Alarm Volume | 8levels setting |
IV Set Management and Flow Rate Calibration | Up to 10 adjustment values for various brands IV set could be stored to make the operation more convenient. |
Anti-reverse Function | Anti-reverse function on motor to prevent upstream |
Self-check | Self-check when start up and during infusion |
Audio and Visual Alarms | Air-in-line, Door open, Occlusion, Infusion finished, No operation, Battery low, Pump error, Motor reverse, Pressure abnormity, Sound abnormity, AC power lost. |
Display | Color LCD with high brightness |
Power Source | AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA |
Battery | Rechargeable Ni-MH battery: DC12V, 2000mAh, life: ≥ 4 hours at 25mL/h |
Operating Conditions | Temperature: 5°C ~ 40°C Relative humidity: 20% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Storage Conditions | Temperature: -20°C ~ 55°C Relative humidity: 10% ~ 90% Atmospheric pressure: 70 ~ 106KPa |
Classification | Class II, CF, IPX4 |
Dimensions and N.W. | 150mmx 150mm x 235mm; 2.3kg |
Application | Intravenous infusion for various liquid medicines, such as vascular expansion solution, anticancer drug, accelerative procreates drug, liquid nutrient, etc. Applicable for micro and rapid infusion. |