RF Conducted Immunity Test Systems (61000-4-6)

After-sales Service: Avaialbe
Warranty: Avaialbe
Product Name: RF Conducted Immunity Test Systems (61000-4-6)

Products Details

Basic Info.

Model NO.
CIT6000-25
Standard
IEC 61000-4-6
Signal Source
4kHz~400MHz
Power Amplifier
100kHz~230MHz, Power 85W
Power Meter
4kHz~500MHz
Calibration Certificate
Available
Certificate
ISO 9001, CE
Custom Service
Available
Transport Package
Wooden Package
Trademark
EMCSOSIN
Origin
Shanghai

Product Description

RF Conducted Immunity Test Systems (61000-4-6)

Product Information

CIT6000-25 RF (Radio-frequency) conducted immunity test equipment adopts an integrated design, with built-in signal generator, RF power amplifier, and RF power meter. The device supports external power amplifiers, which has stronger system scalability. CIT6000-25 is equipped with a built-in industrial computer, with a dedicated RF conducted immunity testing software, which can perform fully automatic calibration and testing, greatly improving system scalability and testing efficiency. It supports users to view, save, and print test results on their own.

Features
  • Built-in integrated signal generator, RF power amplifier and RF power meter
  • Signal source: 4kHz~400MHz
  • Power amplifier: 100kHz~230MHz, power 85W
  • Power meter: 4kHz~500MHz
  • Support external power amplifier
  • 10.1" large color screen, 1280(H)×800(V) screen resolution
  • Standard RF conducted immunity test software

Technical Data
 
Model CIT6000-25
Name RF conducted immunity test equipment
Standard IEC 61000-4-6, GB/T 17626.6
Power supply AC 220V±10%, 50/60Hz
Ambient temperature 0ºC~+40ºC
Relative humidity ≤90%
Signal generator
Frequency range 4kHz~400MHz
Frequency resolution 0.1Hz
Frequency temperature stability ±0.5ppm
Harmonic distortion ≤-25dBc
Nonharmonic distortion ≤-45dBc
Output power range -60dBm~+10dBm
Power accuracy ±1.0dB
Power resolution 0.01dB
Internal modulation source Sine wave: 0.1Hz~1.25MHz
Square wave: 0.1Hz~100kHz
Triangle wave/sawtooth wave: 0.1Hz~100kHz
Amplitude Modulation depth: 1~99%
Modulation rate: 20Hz~1.25MHz
FM Max frequency deviation: 10MHz
Modulation rate: 20Hz~1.25MHz
Phase Phase modulation: 0°~360°
Modulation rate: 20Hz~1.25MHz
Pulse modulation Pulse period: 200ns~160s
Pulse width: 100ns~85s
Duty cycle: 1%~99%
Output interface type N(F)
Output standing wave ratio <1.5:1
Power meter
Frequency range 4kHz~500MHz;
Options support extended frequency range, max frequency can be extended to 1GHz or 6GHz
Dynamic range -60dBm~+20dBm
Power accuracy ±0.2dB
Input VSWR <1.15:1
Input interface type Front: BNC (F)
Rear: SMA (F)
Power amplifier
Frequency range 100kHz~230MHz
Power gain 50dB
Max output power 100W
1dB compression point output power 85W
Input impedance 50Ω, VSWR<1.5:1
Output impedance 50Ω
Output interface type N(F)
Standard accessories
6dB attenuator, 50Ω terminal load, CDN*1,CDN calibration kit*2, RF test cable
Power cable, grounding wire, manual, test line, fuse*2

Company Information
Shanghai EMCSOSIN/ Shanghai SHINEGIL is a professional supplier that specialized in R&D, manufacturing and selling EMC testing equipments. We've been committed to the electromagnetic compatibility (EMC) testing field, offering professional equipments which are widely used in electronics manufacturing, electric power, new energy, telecommunication, aerospace, teaching, renewable energy, automotive and other industries, as well as scientific research. We provide customers with professional EMC system integration solutions. We are also ISO9001 certified.
Our self-developed electrostatic discharge (ESD) generator, electrical fast transient (EFT)/ burst generator, lightning surge generator, voltage dips simulator and so on enjoy a high reputation both at home and abroad. To provide best service and create value for customers is our eternal pursuit!
RF Conducted Immunity Test Systems (61000-4-6)

Manufactory and Office
RF Conducted Immunity Test Systems (61000-4-6)
RF Conducted Immunity Test Systems (61000-4-6)
RF Conducted Immunity Test Systems (61000-4-6)

Qualification and Certificate
RF Conducted Immunity Test Systems (61000-4-6)

Our Client
RF Conducted Immunity Test Systems (61000-4-6)

Packing
RF Conducted Immunity Test Systems (61000-4-6)

Our Service
1) Our product engineers and application technicians answer various questions and deal with emergencies in real time.
2) Professional engineers will guide the installation & debugging of equipments so that users can operate devices independently and solve problems in daily test.
3) Free warranty for 1 year and lifelong maintenance.
4) Free upgrade service for the users is available as per international standard updates requirements.

If you have any further query on EMC/ EMI testing equipments, electrical fast transient/ burst generators, pls do not hesitate to contact us at linda<at>shinegil<dot>com
 
We look forward to working with you sincerely!

Contact us

Please feel free to give your inquiry in the form below We will reply you in 24 hours